For High Resolution Water Level Measurements
The DCX-38 VG is an autonomous instrument for recording water level with a high resolution and full scale ranges as low as 0,5 mWC / 50 mbar. It features a rugged, gold-plated ceramic diaphragm for outstanding long-term stability and stainless steel housing with user serviceable battery for long service life.
The data collector can individually be adjusted to the user's needs and offers the following
advantages:
- Recording of the pressure and temperature
- Simple and well structured configuration- and read-out software (Logger 4.x) for PC or PDA
- Combination of event-controlled recording and interval recording prevents unnecessary data being recorded (i.e. only measuring the pressure changes)
- Installation data (and comments) of the measuring station can be stored in the device
Configuration / Data Retrieval:
For data readout, the serial interface is accessed via the data port located on one end of the cable, enabling on site data retrieval. Using either the K-104A or K-103A converter cable, the DCX-38 VG connects easily to a laptop or PC via USB or serial port. Used in conjunction with Logger 4.X software, the DCX-38 VG is configured and data retrieved either in the lab or the field.
Pressure Ranges | 0…50 mbar 0…100 mbar 0…300 mbar |
Error Band | max. 0,2 %FS |
Resolution |
max. 0,002 %FS |
Stability | FS = 100 mbar: ± 0,1 %FS FS = 100 mbar: ± 0,1 mbar |